Automatic Ignition Tester

Determination of the ignition temperature of flammable liquids and gases for the assessment of the hazard potential of these substances

Properties

  • Ignition Temperature of n-Heptan in < 3 h
  • Touchscreen guides through entire test procedure
  • automatic blow-out device for the test piston
  • Temperature Range +65 °C to +650 °C

Datasheet as PDF Questions?

Description

The device is equipped with a touch screen for setting and displaying the experimental parameters and guides the user step by step through the measurement sequence (Predeterminition, Method S and Manual Determinition).
By means of an active temperature control, consisting of ceramic heating elements and automatic air cooling, it is possible to realize very short test times. For a complete measurement (Predetermination ~ 1:00 h, Method S ~ 1:35 h) of the ignition value of
n-heptane less than 3 hours are required. The device has very good reproducibility by automatic detection of the ignition temperature and indication of the ignition delay time. It also has a mounted automatic blow-out device for the test piston.

Temperature Range

+65 °C to +650 °C

Test Duration

under 3 h for n-Heptan

Test Methods

Predetermination, Method S, Manual Determination

Dimesions Control Box (W x D x H)

approx. 76 x 62 x 24 cm

Dimensions Test Unit (W x D x H)

approx. 46 x 36 x 73 (100 hochgefahren) cm

Weight

approx. 25 kg (Control Box), approx. 49 kg (Test Unit)

Mains

3 ~ 400 V AC, 50 Hz, 16 A CEE Plug

DIN EN 14522

DIN 51794 (in Anlehnung / similar to)

More products

Setaflash Series 3 High Temperature

Flash point tester, semi-automatic, closed cup, RT up to 300 °C

Read more

Setaflash Series 8 High Temperature

Flashpoint tester, automatic, RT up to 300 °C

Read more

Pensky Martens Flashpoint, automatic

Automatic determination of the flash point of petrochemical products according to the Pensky Martens test method in a closed cup

Read more